Polyethlene nanoplastic (PE-NPS)
2026-07-09
Magnetic Bead-Based HPV Nucleic Acid Extraction Kit
2026-07-1496 Channe Automated Nucleic Acid Extractor
Product Number: EX-9601
Shipping and Storage
Packaged nucleic acid extractors shall be stored in a well-ventilated environment at 10℃ ~ 30℃ with relative humidity ≤70% and no corrosive gas.
The maximum storage period for packaged instruments is 1 year. Instruments stored for more than one year must be unpacked and inspected; only qualified instruments may re-enter circulation.
Instrument Structure


The experimental chamber contains the following components as shown in Figure:
- Magnetic rod rack and magnetic sleeves
- Drip-proof plate
- Lysis position, numbered 1 from left to right
- Magnetic bead position, numbered 2 from left to right
- Washing position 1, numbered 3 from left to right
- Washing position 2, numbered 4 from left to right
- Elution position, numbered 5 from left to right
Application
Widely applied in general scientific research, genomics, disease control systems, food safety, forensic science and other fields. Simply load samples and magnetic bead-based fully automatic nucleic acid extraction reagents into the 96-well deep-well plate, then select or edit an appropriate program to execute. Matching different types of magnetic bead nucleic acid extraction reagents enables rapid extraction of DNA and RNA from animal and plant tissues, blood, body fluids, criminal forensic specimens and other samples.
Operating Environment
- Ambient temperature: 10 ℃ ~ 30 ℃
- Relative humidity: ≤70%
- Atmospheric pressure: 86 kPa ~ 106 kPa
- No strong vibration or corrosive gas in the surrounding area
- Avoid direct sunlight or other cold/heat sources
- Keep away from strong electromagnetic field interference
- Power supply voltage: 220 V, 50 Hz
Technical Parameters
| Item | EX-9601 |
| Maximum sample throughput | 96 samples per run |
| Magnetic rods | 96 independent magnetic rods |
| Sample volume range | 20~1000μL |
| Magnetic bead recovery efficiency | >95% |
| Optimal magnetic bead size | 0.2~1.0μm |
| Overall dimensions (mm) | 800*500*500 (L*W*H) |
| Service life | 5 years |
Performance
- Sample protection: Equipped with power-on self-test and power failure protection functions to minimize sample loss during operation.
- Modular design: Adopts modular structure with self-developed core components, delivering higher efficiency, lower failure rate and stable instrument operation.
- User-friendly interface: Visualized programs with precise control for simple, easy-to-learn operation.
- Independent programmable control: Supports free program editing per user requirements.
- Open platform: Compatible with multiple biological nano-magnetic bead-based nucleic acid extraction methods.
Main Structure
The nucleic acid extractor mainly consists of heating module assembly, magnetic separation module assembly, industrial control module assembly, base frame module assembly and outer casing assembly.